Kon, Ryan; Jain, Ameet K.; Fichtinger, Gabor
Hidden seed reconstruction from C-arm images in brachytherapy Conference
Procedings 3rd IEEE International Symposium on Biomedical Imaging: Nano to Macro, 2006.
@conference{Kon2006,
title = {Hidden seed reconstruction from C-arm images in brachytherapy},
author = {Ryan Kon and Ameet K. Jain and Gabor Fichtinger},
url = {https://labs.cs.queensu.ca/perklab/wp-content/uploads/sites/3/2024/02/Kon2006.pdf},
doi = {10 1109/ISBI 2006 1624969},
year = {2006},
date = {2006-04-01},
urldate = {2006-04-01},
booktitle = {Procedings 3rd IEEE International Symposium on Biomedical Imaging: Nano to Macro},
pages = {526–529},
keywords = {},
pubstate = {published},
tppubtype = {conference}
}
Jain, Ameet K.; Kon, Ryan; Zhou, Yu; Fichtinger, Gabor
C-arm calibration-is it really necessary? Journal Article
In: Medical image computing and computer-assisted intervention (MICCAI), vol. 8, no. Pt 1, pp. 639–646, 2005.
@article{Jain2005,
title = {C-arm calibration-is it really necessary?},
author = {Ameet K. Jain and Ryan Kon and Yu Zhou and Gabor Fichtinger},
url = {https://labs.cs.queensu.ca/perklab/wp-content/uploads/sites/3/2024/02/Jain2005c.pdf},
year = {2005},
date = {2005-01-01},
urldate = {2005-01-01},
journal = {Medical image computing and computer-assisted intervention (MICCAI)},
volume = {8},
number = {Pt 1},
pages = {639–646},
abstract = {<p>C-arm fluoroscopy is modelled as a perspective projection, the parameters of which are estimated through a calibration procedure It has been universally accepted that precise intra-procedural calibration is a prerequisite for accurate quantitative C-arm fluoroscopy guidance Calibration, however, significantly adds to system complexity, which is a major impediment to clinical practice We challenge the status quo by questioning the assumption that precise intra-procedural calibration is really necessary We derived theoretical bounds for the sensitivity of 3D measurements to mis-calibration Experimental results corroborated the theory in that mis-calibration in the focal spot by as much as 50 mm still allows for tracking with an accuracy of 0 5 mm in translation, 0 65 degrees in rotation, and such mis-calibration does not impose any additional error on the reconstruction of small objects</p>},
keywords = {},
pubstate = {published},
tppubtype = {article}
}
<p>C-arm fluoroscopy is modelled as a perspective projection, the parameters of which are estimated through a calibration procedure It has been universally accepted that precise intra-procedural calibration is a prerequisite for accurate quantitative C-arm fluoroscopy guidance Calibration, however, significantly adds to system complexity, which is a major impediment to clinical practice We challenge the status quo by questioning the assumption that precise intra-procedural calibration is really necessary We derived theoretical bounds for the sensitivity of 3D measurements to mis-calibration Experimental results corroborated the theory in that mis-calibration in the focal spot by as much as 50 mm still allows for tracking with an accuracy of 0 5 mm in translation, 0 65 degrees in rotation, and such mis-calibration does not impose any additional error on the reconstruction of small objects</p>
Kon, Ryan; Leven, Joshua; Kothapalli, Kishore; Boctor, Emad; Fichtinger, Gabor; Hager, Gregory D.; Taylor, Russell H.
CIS-UltraCal: an open-source ultrasound calibration toolkit Conference
Medical Imaging 2005: Ultrasonic Imaging and Signal Processing, vol. 5750, no. 1, SPIE SPIE, 2005.
@conference{Kon2005,
title = {CIS-UltraCal: an open-source ultrasound calibration toolkit},
author = {Ryan Kon and Joshua Leven and Kishore Kothapalli and Emad Boctor and Gabor Fichtinger and Gregory D. Hager and Russell H. Taylor},
url = {http://link aip org/link/?PSI/5750/516/1
https://labs.cs.queensu.ca/perklab/wp-content/uploads/sites/3/2024/02/Kon2005.pdf},
doi = {10 1117/12 596929},
year = {2005},
date = {2005-01-01},
urldate = {2005-01-01},
booktitle = {Medical Imaging 2005: Ultrasonic Imaging and Signal Processing},
volume = {5750},
number = {1},
pages = {516-523},
publisher = {SPIE},
organization = {SPIE},
keywords = {},
pubstate = {published},
tppubtype = {conference}
}